Feature |
Meso 200A |
Micro 200B |
Micro 200C |
Adsorbed Gas |
Non corrosive gases, such as N2, Ar, Kr, H2, O2, CO2,
CO, NH3, CH4, etc.
|
Pressure Sensor at
Analysis Station
|
1000 torr, 2 |
1000 torr, 2;
10 torr, 1 (optional 2);
1 torr(0.1torr can be selected), optional 1
|
1000 torr, 2;
10 torr, 2;
1 torr (0.1torr can be selected), 2
|
Accuracy: ±0.15% (F.S.) |
Pressure Sensor at
P0 Station
|
1000 torr,2
(Accuracy: ±0.15% (F.S.)
|
Degas System |
Two ectopic sample preparation ports;
(Analysis station supports in situ sample preparation)
|
Degas Temperature |
Ambient to 400 ℃. Free to set up target temperature. |
Cold Trap |
1 |
Vacuum Pump |
Two-stage rotary vane mechanical vacuum pump,
the ultimate pressure is
6.7*10-2 Pa
|
Two-stage rotary vane mechanical vacuum pump, the ultimate pressure is
6.7*10-2 Pa, optional turbo molecular pump
|
Turbo molecular pump (ultimate pressure 10-8 Pa) and front mechanical vacuum
pump (ultimate pressure is 6.7*10-2 Pa)
|
Analysis Port |
Samples on the 2 analysis bits can be tested alternately (including P0 test). |
Test Principle |
Gas adsorption by static-volumetric analysis |
Measurement Range of Specific Surface Area |
0.0005 m2/g to the infinity; Standard sample repeat
accuracy is less than ± 1.0%
|
0.0001m2/g to the infinity; Standard sample repeat accuracy is less than ± 1.0%
|
0.0001m2/g to the infinity; Standard sample repeat accuracy is less than ± 1.0%
|
Test Range of Pore Diameter |
0.35 nm-500 nm;
Pore dimension repeatability is less than 0.2 nm in the accurate analysis of materials with pore
size greater than 2 nm
|
0.35 nm-500 nm;
Pore dimension repeatability is less than 0.02 nm in the accurate analysis of micropore 0.35
nm-2 nm
|
0.35 nm-500 nm;
Pore dimension repeatability is less than 0.02 nm in the
accurate analysis of micropore 0.35 nm-2 nm
|
Minimum Pore Volume |
0.0001 cm3/g |
Range of Relative
Pressure P/P0
|
10-5-0.998 |
10-6/10-8-0.998 |
10-8-0.998 |
Overall Dimension |
Depth: 870 mm; width: 570 mm; height: 890 mm; weight: 80 Kg -90 Kg |
Ambient Temperature |
15-40 oC |
Related Humidity |
30%-60% |
Electrical Supply |